Medicine-Engineering Interdisciplinary

Experimental Investigation of Possibility of Simultaneously Monitoring Lung Perfusion/Cardiomotility and Ventilation via Thoracic Impedance Measurement

  • BAI Zixuan 1(白子轩) ,
  • MA Yixin1,2?(马艺馨) ,
  • KONG Zhibin3(孔志斌) ,
  • XUE Shan4 (薛珊)
Expand
  • (1. School of Electronic Information and Electrical Engineering, Shanghai Jiao Tong University, Shanghai 200240, China; 2. Shanghai Engineering Research Center for Intelligent Diagnosis and Treatment Instrument, Shanghai 200240, China; 3. The Sixth People’s Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200235, China; 4. Renji Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai 200127, China)

Accepted date: 2023-04-10

  Online published: 2025-01-28

Abstract

Impedance pneumography has a significant advantage for continuous and noninvasive monitoring of respiration, compared with conventional flowmeter-based ventilation measurement technologies. While thoracic impedance is sensitive to pulmonary ventilation, it is also sensitive to physiological activities such as blood flow and cardiomotility, in addition, body movement/posture. This paper explores the possibility of simultaneously monitoring pulmonary ventilation, blood circulation and cardiomotility by bioimpedance measurement. Respiratory, blood perfusion and cardiomotility signals are extracted using the wavelet method from thoracic impedance data measured in breath-holding and tidal breathing statuses, to investigate signal strength and their dependency. This research provides a foundation for the development of bedside devices to monitor various physiological activities.

Cite this article

BAI Zixuan 1(白子轩) , MA Yixin1,2?(马艺馨) , KONG Zhibin3(孔志斌) , XUE Shan4 (薛珊) . Experimental Investigation of Possibility of Simultaneously Monitoring Lung Perfusion/Cardiomotility and Ventilation via Thoracic Impedance Measurement[J]. Journal of Shanghai Jiaotong University(Science), 2025 , 30(1) : 81 -90 . DOI: 10.1007/s12204-023-2639-2

References

[1] GUPTA Y S, SHAH S S, AHIRE C K, et al. Body plethysmography in chronic obstructive pulmonary disease patients: A cross-sectional study [J]. Lung India: Official Organ of Indian Chest Society, 2018, 35(2): 127-131.
[2] MLYNCZAK M. Physiological measurements of cardiorespiratory activity with impedance pneumography [M]//Bioimpedance and spectroscopy. Amsterdam: Elsevier, 2021: 337-394.
[3] ALLISON R D, HOLMES E L, NYBOER J. Volumetric dynamics of respiration as measured by electrical impedance plethysmography [J]. Journal of Applied Physiology, 1964, 19: 166-173.
[4] SEPPA V P, HYTTINEN J, UITTO M, et al. Novel electrode configuration for highly linear impedance pneumography [J]. Biomedizinische Technik Biomedical Engineering, 2013, 58(1): 35-38.
[5] QIU C K, WU F, HAN W Q, et al. A wearable bioimpedance chest patch for real-time ambulatory respiratory monitoring [J]. IEEE Transactions on BioMedical Engineering, 2022, 69(9): 2970-2981.
[6] MLYNCZAK M, NIEWIADOMSKI W, ZYLINSKIM, et al. Assessment of calibration methods on impedance pneumography accuracy [J]. Biomedizinische Technik Biomedical Engineering, 2016, 61(6): 587-593.
[7] MLYNCZAK M, KO ′ LODZIEJCZYK A, KRYSZTOFIAK H, et al. Cardiorespiratory profiling during simulated lunar mission using impedance pneumography [J]. Biomedical Signal Processing and Control, 2019, 51: 216-221.
[8] CHARLTON P H, BONNICI T, TARASSENKO L, et al. An impedance pneumography signal quality index: Design, assessment and application to respiratory rate monitoring [J]. Biomedical Signal Processing and Control, 2021, 65: 102339.
[9] MOHR M A, FAIRCHILD K D, PATEL M, et al. Quantification of periodic breathing in premature infants [J]. Physiological Measurement, 2015, 36(7):1415-1427.
[10] AQUEVEQUE P, GOMEZ B, MONSALVE E, et al. Simple wireless impedance pneumography system for unobtrusive sensing of respiration [J]. Sensors, 2020,20(18): 5228.
[11] NYBOER J. Electrical impedance plethysmograph [M]. Springfield: Thomas Publishers, 1970.
[12] BAKER L E. Principles of the impedance technique [J]. IEEE Engineering in Medicine and Biology Magazine, 1989, 8(1): 11-15.
[13] LIU E. Research on intracardiac catheter positioning technology based on electric field characteristics [D]. Shanghai: Shanghai Jiao Tong University, 2019 (in Chinese).
[14] ANAND G, LOWE A. Investigating electrical impedance spectroscopy for estimating blood flow-induced variations in human forearm [J]. Sensors,2020, 20(18): 5333.
[15] MELNIKOV A A, POPOV S G, NIKOLAEV D V, et al. The measurement of peripheral blood volume reactions to tilt test by the electrical impedance technique after exercise in athletes [J]. Journal of Physics: Conference Series, 2013, 434: 012055.
[16] SCUDDER M R, JENNINGS J R, DUPONT C M, et al. Dual impedance cardiography: An inexpensive and reliable method to assess arterial stiffness [J]. Psychophysiology, 2021, 58(7): e13772.
[17] LI K Y, CHEN W H, ZHENG L R, et al. A method to detect heart rate based on electrical bio-impedance [J]. ITM Web of Conferences, 2016, 7: 09014.
[18] HAFID A, BENOUAR S, KEDIR-TALHA M, et al. Simultaneous recording of ICG and ECG using Z-RPI device with minimum number of electrodes [J]. Journal of Sensors, 2018, 2018: 1-7.
[19] LEE H, RUSIN C G, LAKE D E, et al. A new algorithm for detecting central apnea in neonates [J]. Physiological Measurement, 2012, 33(1): 1-17.
[20] LI H, XU H, MA Y X, et al. Thoracic impedance measurement for lung function evaluation [C]//2017 IEEE International Conference on Imaging Systems and Techniques. Beijing: IEEE, 2017: 1-5.
[21] BROWN B H, BARBER D C, MORICE A H, et al. Cardiac and respiratory related electrical impedance changes in the human thorax [J]. IEEE Transactions on Biomedical Engineering, 1994, 41(8): 729-734.
[22] MLY NCZAK M, NIEWIADOMSKI W, ZYLI NSKI M, et al. Verification of the respiratory parameters derived from impedance pneumography during normal and deep breathing in three body postures [M]//6th European conference of the international federation for medical and biological engineering. Cham: Springer, 2015: 881-884.
[23] YANG L, QU S Y, ZHANG Y W, et al. Removing clinical motion artifacts during ventilation monitoring with electrical impedance tomography: Introduction of methodology and validation with simulation and patient data [J]. Frontiers in Medicine, 2022, 9: 817590.
[24] BLANCO-ALMAZAN D, GROENENDAAL W, CATTHOOR F, et al. Chest movement and respiratory volume both contribute to thoracic bioimpedance during loaded breathing [J]. Scientific Reports, 2019, 9: 20232.
[25] HUA P, WOO E J, WEBSTER J G, et al. Using compound electrodes in electrical impedance tomography [J]. IEEE Transactions on Biomedical Engineering, 1993, 40(1): 29-34.
[26] SHI F, QIAO X, SONG G, et al. Clinical observation of bipap noninvasive ventilator therapy for copd complicated with type II respiratory failure [J]. China Medical Device Information, 2020, 26(8): 138-140 (in Chinese).
[27] MALLAT S G, PEYRE G. A wavelet tour of signal ′ processing: The sparse way [M]. 3rd ed. Amsterdam: Elsevier/Academic Press, 2009.
[28] GUYTON A C, HALL J E. Textbook of medical physiology [M]. 11th ed. Philadelphia: Elsevier Saunders,2006: 106.
Outlines

/