Abstract: Correlated photon is exploited as a tool for absolute quantum efficiency calibration of photon detectors
working in the photon-counting regime. We extend the method from photon-counting regime to the absolute
calibration of analog regime photon detectors. A revised calibration module is proposed based on balance detection
and fluctuation restrictions conditions applied to analog photon detectors. An absolute calibration experimental
system is established. This method shows a relative standard uncertainty of 2.1% and a relative deviation of 1%
compared with those of the typical photomultiplier quantum efficiency.
GAO Dongyang1,2 (高冬阳), XIA Maopeng1 (夏茂鹏), LI Jianjun1 (李健军),HU Youbo1 (胡友勃), ZHENG Xiaobing1,2* (郑小兵)
. Absolute Quantum Efficiency Calibration of Analog Detectors with Twin Photon Current[J]. Journal of Shanghai Jiaotong University(Science), 2017
, 22(4)
: 411
-416
.
DOI: 10.1007/s12204-017-1855-z
[1] KLYSHKO D N. Use of two-photon light for absolute calibration of photoelectric detectors [J]. Soviet Journal Quantum Electronics, 1980, 10(9): 1112-1116.
[2] POLYAKOV S V, MIGDALL A L. High accuracy verification of a correlated photon-based method for determining photon-counting detection efficiency [J]. Optics Express, 2007, 15(4): 1390-1404.
[3] BRIDA G, CASTELLETTO S, NOVRO C, et al. Measurement of the quantum efficiency of photodetectors by parametric fluorescence [J]. Metrologia, 1998, 35(4):379-401.
[4] CHEUNG J Y, CHUNNILALL C J, Thomas P J, et al. Photon-counting measurement challenges [J]. Proceeding of SPIE, 2007, 6583: 65830G1-65830G12.
[5] GHAZI-BELLOUATI A, RAZETA A, BASTIE J, et al. Detector calibration at INM using a correlated photons source [J]. European Physical Journal Applied Physics, 2006, 35: 211-216.
[6] CHANG J T, WU L A. Absolute self-calibration of the quantum efficiency of single-photon detectors [J]. Acta Physica Sinica, 2003, 52(5): 1132-1135 (in Chinese).
[7] LV L, ZHANG Y C, LIN Y D. Research on absolute calibration of photodetector quantum-efficiency using entangled photons [J]. Acta Optica Sinica, 2012, 32(1):119-125 (in Chinese).
[8] LI J J, ZHENG X B, LU Y J, et al. Quantum efficiency calibration of opto-electronic detector by means of correlated photons method [J]. Chinese Optics Letter,2008, 6(7): 472-475.
[9] ZHAO K, LIU C M, CHENG H D, et al. Measurement of InGaAs single photon detector quantum efficiency at 1550 nm [J]. Proceeding of SPIE, 2015, 9449: 94492Y.
[10] BRIDA G, DEGIOVANNI I P, GENOVESE M, et al.Detection of multmode spatial correlation in PDC and application to the absolute calibration of a CCD camera[J]. Optics Express, 2010, 18(20): 20572.
[11] MEDA A, RUO-BERCHERA I, DEGIOVANNI I P,et al. Absolute calibration of a charge-coupled device camera with twin beams [J]. Applied Physics Letters,2014, 105: 101113.
[12] BRIDA G, GENOVESE M, RVO-BERCHERA I. Possibility of absolute calibration of analog detectors by using parametric down-conversion: A systematic study[J]. Journal of the Optical Society of America B, 2006,23(10): 2158-2193.