Computational Modeling and Experimental Validation of MRI-Induced RF Heating for a Transcranial Puncture Needle at 3 T

3T磁共振成像下经颅穿刺针射频致热的计算建模与实验验证

高中1,国然2,李铖铃2,高安柱3,朱为然4,孔继烈1   

  1. 1. Department of Chemistry, Fudan University, Shanghai 200433, China; 2. Shanghai Institute of Medical Device Testing, Shanghai 201318, China; 3. School of Automation and Intelligent Sensing; Institute of Medical Robotics, Shanghai Jiao Tong University, Shanghai 200240, China; 4. SceneRay Co., Ltd., Suzhou 215128, Jiangsu, China
  2. 1. 复旦大学 化学系,上海200433;2. 上海市医疗器械检验研究院,上海201318;3. 上海交通大学 自动化与感知学院;医疗机器人研究院,上海 200240;4. 景昱医疗科技(苏州)股份有限公司,江苏苏州215128
  • Received:2026-01-12 Revised:2026-04-12 Accepted:2026-05-08 Online:2026-06-28 Published:2026-06-11

Abstract: This study presents a computational and experimental framework for assessing magnetic resonance imaging (MRI)-induced radiofrequency (RF) heating risks for a transcranial puncture needle at 3T MRI. Due to the one-dimensional structure of the puncture needle, as well as its clinical interventional use, the transfer function method was used to evaluate the needle’s electromagnetic model. Multi-variable computational simulations encompassing over 100 000 tests were performed using five anatomical virtual human models, including various coil configurations and clinical implantation trajectories. The in vivo RF heating was predicted by convolving the simulated tangential electric field along each path with the measured transfer function, scaled by an experimentally validated factor. The maximum temperature rise of the puncture needle in the human body does not exceed 6 °C after a 15-min scan under a given power limit exposure under MRI. This integrated approach provides a robust safety assessment for transcranial MRI-guided intervention.

Key words: radiofrequency (RF) heating, transcranial puncture needle, magnetic resonance imaging (MRI)

摘要: 本研究提出一种计算与实验相结合的框架,用于评估3 T磁共振成像(MRI)下经颅穿刺针的射频(RF)致热风险。鉴于穿刺针的一维结构及其临床介入用途,采用传递函数法建立穿刺针的电磁模型。基于五个解剖虚拟人体模型,结合不同线圈配置和临床植入轨迹,开展了超过10万次多变量计算仿真。通过将沿各路径的切向电场仿真值与实测传递函数进行卷积,并乘以经实验验证的校正因子,对体内射频致热进行预测。结果表明:在给定功率限值暴露条件下,MRI扫描15分钟后,穿刺针在人体内的最高温升不超过6 ℃。该综合方法为MRI引导下的经颅介入操作提供了可靠的安全性评估。

关键词: 射频致热,经颅穿刺针,磁共振成像